Loading... Please wait...Currency Displayed in
| DESCRIPTION | ||||||||||||||||||||
|
High Pressure Purifier Moisture and Oil Removal, Brass Housing Model SG6140 Model SG6140 purifier is designed to remove water and/or oil from gas or liquid streams. In the laboratory, it protects delicate or sensitive instruments, prolongs service life and provides more accurate, reproducible test results. This purifier may be used with noncorrosive gases compatible with brass and Viton® at pressures up to 3000 psig. Three different purifying elements are available (see Tables I and II). These elements can be replaced without removing the purifier from the process line. Note: These purifiers are not suitable for use with oxygen. Model SG6140 cannot be used as a particulate filter. For particulate filters see pages 190–192. Specifications Maximum Operating Pressure: 3000 psig Pressure Drop: 1.0 psi at 56 slpm Air 2.4 psi at 85 slpm Air Dew Point Obtainable (4A and 13X Molecular Sieve): -100°F Inlet & Outlet Connections: 1⁄4" NPT female Dimensions: Height: 5 1⁄16" Diameter (Widest Point): 2" Mounting Holes (2): 1⁄4" -20UNC x 3⁄8" deep Approximate Weight: 3 lbs Materials of Construction Purifier: Body and Cap: Brass O-Ring Seal: Viton® Purifier Element: Housing: Electrolytic Tin-Plated Cold-Rolled Steel Retainer: Polyester Felt Backed by Type 316 SS Screen Desiccant: 4A, 13X or Activated Charcoal as Ordered |
||||||||||||||||||||
|
TYPICAL LEAD-TIME 1-10 DAYS |
||||||||||||||||||||